日韩欧美一区二区三区免费观看_精品国产欧美一区二区_波多野结衣乱码中文字幕_最新无码国产在线视频2021

技術(shù)中心

CSB-E09502h人降鈣素原(PCT)ELISA試劑盒說明書

2011年12月21日 16:02:08人氣:249來源:廈門慧嘉生物科技有限公司

資料類型doc文件資料大小98380
下載次數(shù)124資料圖片 【點擊查看】
上 傳 人廈門慧嘉生物科技有限公司 需要積分0
關(guān) 鍵 詞人降鈣素原,PCT,ELISA試劑盒
【資料簡介】

 

 Human Procalcitonin (PCT) ELISA Kit
Catalog No. CSB-E09502h
(96 tests)
This immunoassay kit allows for the in vitro quantitative determination of human PCT concentrations in serum, plasma and other biological fluids.
Expiration date six months from the date of manufacture
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
 
廈門慧嘉生物長期經(jīng)營ELISA試劑盒及Santa/Abcam抗體、Prospec細胞因子、Sigma/Amresco/Qiagen、Axygen耗材等生物試劑產(chǎn)品。誠信經(jīng)營,價格實惠,服務(wù)周到,質(zhì)量有保證。歡迎廣告老師來詢!:   :  1048735792 或登陸/download(向客服人員索取原版說明書)
 
INTRODUCTION
Procalcitonin (PCT) is a precursor of the hormone calcitonin, which is involved with calcium homeostasis, and is produced by the C-cells of the thyroid gland. It is there that procalcitonin is cleaved into calcitonin, katacalcin and a protein residue.
Measurement of procalcitonin can be used as a marker of severe sepsis and generally grades well with the degree of sepsis, although levels of procalcitonin in the blood are very low. PCT has the greatest sensitivity (85%) and specificity (91%) for differentiating patients with SIRS from those with sepsis, when compared with IL-2, IL-6, IL-8, CRP and TNF-alpha. Evidence is emerging that procalcitonin levels can reduce unnecessary antibiotic prescribing to people with lower respiratory tract infections. Currently though, procalcitonin assay is not used outside the context of clinical trials.
PRINCIPLE OF THE ASSAY
The microtiter plate provided in this kit has been pre-coated with an antibody specific to PCT. Standards or samples are then added to the appropriate microtiter plate wells with a Horseradish Peroxidase (HRP) -conjugated antibody preparation specific for PCT and incubated. Then substrate solutions are added to each well. The enzyme-substrate reaction is terminated by the addition of a sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450 nm ±2 nm. The concentration of PCT in the samples is then determined by comparing the O.D. of the samples to the standard curve.
DETECTION RANGE
125 pg/ml-4000 pg/ml. The standard curve concentrations used for the ELISA’s were 4000 pg/ml, 2000 pg/ml, 875 pg/ml, 375 pg/ml, 125pg/ml.
SPECIFICITY
This assay recognizes human PCT. No significant cross-reactivity or interference was observed.
SENSITIVITY
The minimum detectable dose of human PCT is typically less than 62.5 pg/ml.
The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as the lowest protein concentration that could be differentiated from zero.

MATERIALS PROVIDED
 
Reagent
Quantity
 
Assay plate
1
 
Standard
5 x 0.5 ml
 
HRP-conjugate
1 x 6 ml
 
Substrate A
1 x 7 ml
 
Substrate B
1 x 7 ml
 
Stop Solution
1 x 7 ml
 

Standard
S1
S2
S3
S4
S5
Concentration (pg/ml)
125
375
875
2000
4000

STORAGE
1.    Unopened test kits should be stored at 2-8 C upon receipt and the microtiter plate should be kept in a sealed bag. The test kit may be used throughout the expiration date of the kit. Refer to the package label for the expiration date.
2.    Opened test kits will remain stable until the expiring date shown, provided it is stored as prescribed above.
3.    A microtiter plate reader with a bandwidth of 10 nm or less and an optical density range of 0-3 OD or greater at 450nm wavelength is acceptable for use in absorbance measurement.
 
OTHER SUPPLIES REQUIRED
1          Microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.
2          Pipettes and pipette tips.
3          Deionized or distilled water.
4          Squirt bottle, manifold dispenser, or automated microplate washer.
 
SAMPLE COLLECTION AND STORAGE
Serum Use a serum separator tube (SST) and allow samples to clot for 30 minutes before centrifugation for 15 minutes at 1000 x g. Remove serum and assay immediay or aliquot and store samples at -20. Centrifuge the sample again after thawing before the assay. Avoid repeated freeze-thaw cycles.
Plasma Collect plasma using citrate, EDTA, or heparin as an anticoagulant. Centrifuge for 15 minutes at 1000 x g within 30 minutes of collection. Assay immediay or aliquot and store samples at -20. Centrifuge the sample again after thawing before the assay. Avoid repeated freeze-thaw cycles.
 
Note: Grossly hemolyzed samples are not suitable for use in this assay.
ASSAY PROCEDURE
Bring all reagents and samples to room temperature before use. It is recommended that all samples, standards, and controls be assayed in duplicate. All the reagents should be added directly to the liquid level in the well. The pipette should avoid contacting the inner wall of the well.
1         Set a Blank well without any solution. Add 50μl of Standard or Sample per well.
2         Add 50μl of HRP-Conjugate to each well (Not to Blank!). Incubate for 1 hour at 37°C.
3         Aspirate each well and wash, repeating the process three times for a total of three washes. Wash by filling each well with ddH2O (200μl) using a squirt bottle, multi-channel pipette, manifold dispenser or autowasher. Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Buffer by aspirating or decanting. Invert the plate and blot it against clean paper towels.
4         Add 50μl of Substrate A and 50μl Substrate B to each well. Incubate for 15 minutes at 37°C. Keeping the plate away from drafts and other temperature fluctuations in the dark.
5         Add 50μl of Stop Solution to each well. If color change does not appear uniform, gently tap the plate to ensure thorough mixing.
6         Determine the optical density of each well within 30 minutes, using a microplate reader set to 450 nm.
 
CALCULATION OF RESULTS
Average the duplicate readings for each standard, control, and sample and subtract the average zero standard optical density. Create a standard curve by reducing the data using computer software capable of generating a four parameter logistic (4-PL) curve-fit. As an alternative, construct a standard curve by plotting the mean absorbance for each standard on the y-axis against the concentration on the x-axis and draw a best fit curve through the points on the graph. The data may be linearized by plotting the log of the PCT concentrations versus the log of the O.D. and the best fit line can be determined by regression analysis. This procedure will produce an adequate but less precise fit of the data. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.
LIMITATIONS OF THE PROCEDURE
1          The kit should not be used beyond the expiration date on the kit label.
2          Do not mix or substitute reagents with those from other lots or sources.
3          If samples generate values higher than the highest standard, dilute the samples and repeat the assay.
4          Any variation in operator, pipetting technique, washing technique, incubation time or temperature, and kit age can cause variation in binding.
5          This assay is designed to eliminate interference by soluble receptors, binding proteins, and other factors present in biological samples. Until all factors have been tested in the Quantikine Immunoassay, the possibility of interference cannot be excluded.
 
TECHNICAL HINTS
1          When mixing or reconstituting protein solutions, always avoid foaming.
2          To avoid cross-contamination, change pipette tips between additions of each standard level, between sample additions, and between reagent additions. Also, use separate reservoirs for each reagent.
3          When using an automated plate washer, adding a 30 second soak period following the addition of wash buffer, and/or rotating the plate 180 degrees between wash steps may improve assay precision.
4          To ensure accurate results, proper adhesion of plate sealers during incubation steps is necessary.
5          Substrate Solution should remain colorless until added to the plate. Keep Substrate Solution protected from light. Substrate Solution should change from colorless to gradations of blue.
6          Stop Solution should be added to the plate in the same order as the Substrate Solution. The color developed in the wells will turn from blue to yellow upon addition of the Stop Solution. Wells that are green in color indicate that the Stop Solution has not mixed thoroughly with the Substrate Solution.

廈門慧嘉生物科技有限公司作者

上一篇:豚鼠血清總補體(CH50)酶聯(lián)免疫(Elisa)試劑盒說明書

下一篇:輪狀病毒抗原(RV-Ag)ELISA試劑盒說明


我要投稿
  • 投稿請發(fā)送郵件至:(郵件標(biāo)題請備注“投稿”)hbzhan@vip.qq.com
  • 聯(lián)系電話0571-87759680
環(huán)保行業(yè)“互聯(lián)網(wǎng)+”服務(wù)平臺
環(huán)保在線APP

功能豐富 實時交流

環(huán)保在線小程序

訂閱獲取更多服務(wù)

微信公眾號

關(guān)注我們

抖音

環(huán)保在線網(wǎng)

抖音號:hbzhan

打開抖音 搜索頁掃一掃

視頻號

環(huán)保在線

公眾號:環(huán)保在線

打開微信掃碼關(guān)注視頻號

快手

環(huán)保在線

快手ID:2537047074

打開快手 掃一掃關(guān)注
意見反饋
主站蜘蛛池模板: 岫岩| 噶尔县| 永丰县| 双牌县| 佛坪县| 灵宝市| 荃湾区| 山阴县| 恩施市| 延吉市| 靖西县| 德州市| 鲁甸县| 图木舒克市| 永登县| 吴旗县| 云阳县| 齐河县| 延寿县| 丹巴县| 东丰县| 华蓥市| 湖北省| 光泽县| 乾安县| 柞水县| 乌拉特中旗| 铜山县| 宣武区| 当雄县| 金阳县| 东莞市| 富顺县| 英山县| 新龙县| 洪雅县| 钟山县| 商都县| 漳浦县| 乌审旗| 铜山县|